• This email address is being protected from spambots. You need JavaScript enabled to view it.
PEM Model 200 Personal Environmental Monitor

PEM Model 200 Personal Environmental Monitor

Research Institute of Biomolecular and Chemical Engineering | Air Chemistry Research Group
Research field: sampling of atmospheric aerosol, mass concentration, size distribution, gas chromatography, elemental analysis, black carbon, toxicity

The personal environmental monitor is a single-stage impactor with an after-filter. The impactor stage removes aerosol particles larger than the cut size (either 2.5 µm or 10 µm aerodynamic diameter), and the smaller particles are collected on the after-filter. Aerosol particles collected on the after-filter are available for gravimetric and/or chemical composition analysis. The PEM can be used to obtain personal exposure data while the wearer goes about his/her normal activities.